Z Pump Model Ammonia Monitor

Pump Model Ammonia Monitor Image

  • Easy to Use

  • Internal Pump

  • Rechargable Battery

  • Short and Long Term Averaging

  • Internal Memory (Data Logging)


The Z-800XP Ammonia Monitor is a portable desktop instrument which measures ammonia concentration in air with a resolution of 0.1 ppm and a range 0-50 ppm. The LCD display outputs the concentration (in ppm) every 10 seconds. The instrument also computes a 15-minute average and a Time Weighted Average (updated hourly for up to 8 hours). These values can be viewed anytime during the measurement procedure. The internal pump can be combined with an extendable probe, for sampling from hard to reach areas.

With the data-logging feature enabled, the meter will store all of the exposure points for each use. The maximum storage capacity is 14,400 data points. This is the equivalent of 5 separate 8-hour uses. The data is easily uploaded to PC using components available within the Microsoft Windows Operating System or the terminal software included with the instrument.



The sensing element of the instrument is an electrochemical cell. The cell is a four-electrode type, which contains a working and an active auxiliary electrode. The signal from the auxiliary electrode is used for temperature compensation. This improves the selectivity of the entire sensor. The sensors response is linear with the concentration of ammonia in air.



PERFORMANCE DATA

Sensor
Nominal Range
Maximum PPM
Resolution
Response Time
    (90% Peak)
Repeatability
Sensitivity Loss
Operating Temperature
Pressure Range
Relative Humidity
Dimensions

Weight
Power Source

Battery Life
Sensor Life
Electrochemical
0-50 ppm
200 ppm
0.1 ppm

<90 Seconds
5% of output
<4% per month
0-40°C
Atmospheric ±10%
15-90% non-condensing
HxDxW
7.5"x5.75"x2.75"
900 gms
Internal Rechargable Battery
or AC Adapter
9 hours (operating time)
2+ years




Some representative examples of the common compounds and the corresponding signals they give are shown below. Care needs to be exercised if using this instrument in the presence of large concentrations of these gases.



CROSS SENSITIVITY DATA

Gas
Actual Concentration
Z-800XP Output
Carbon Monoxide
Hydrogen Sulfide
Sulphur Dioxide
Nitric Oxide
Nitrogen Dioxide
Chlorine
Hydrogen
Hydrogen Cyanide
Hydrogen Chloride
Ethylene
300
15
5
35
5
1
10
5
45
100
0
15
3
7
0
-0.5
0
0.5
0