Z Pump Model Chlorine Monitor

Pump Model Chlorine Monitor Image

  • Easy to Use

  • Internal Pump

  • Rechargeble Battery

  • Short and Long Term Averaging

  • Internal Memory (Data Logging)


The Z-400XP Chlorine Monitor is a portable desktop instrument for monitoring chlorine concentration in air, with a range of 0-20 ppm, and a resolution of 0.01 ppm. The LCD display updates the concentration (in ppm) every 10 seconds. The instrument also computes a 15 minute average and a Time Weighted Average (updated hourly for up to 8 hours). These values can be viewed anytime during the measurement procedure. When combined with the optional extendable probe, the internal pump makes it possible to sample from hard to reach areas.

With the data logging feature enabled, the meter will store all of the exposure points for each use. The maximum storage capacity is 14,400 data points. This is the equivolent of 5 seperate 8 hour uses. The data is easily uploaded to PC using components available within the Microsoft Windows Operating System or the terminal software included with the instrument.



The sensing element of the instrument is an electrochemical cell. The cell is a four-electrode type, which contains a working and an active auxiliary electrode. The signal from the auxiliary electrode is used for temperature compensation. This improves the selectivity of the entire sensor. The sensors response is linear with the concentration of chlorine in air.




PERFORMANCE DATA

Sensor
Nominal Range
Resolution
Response Time
    (90% Peak)
Repeatability
Sensitivity Loss
Operating Temperature
Pressure Range
Relative Humidity
Dimensions

Weight
Power Source

Battery Life
Sensor Life
Electro-chemical
0-50 ppm
0.01 ppm

<150 Seconds
5% of output
<4% per month
-20-40°C
Atmospheric ±10%
15-90% non-condensing
HxDxW
7.5"x5.75"x2.75"
900 gms
Internal Rechargeble Battery
or AC Adapter
9 hours (operating time)
2+ years




Some representative examples of the common compounds and the corresponding signals they give are shown below. For example, the output from the Z-400XP is 0 ppm when exposed to 300 ppm carbon monoxide in air. Care needs to be exercised about using this instrument in the presence of large concentrations of these gases.



CROSS SENSITIVITY DATA

Gas
Actual Concentration
Z-400XP Output
Carbon Monoxide
Hydrogen Sulfide
Sulfur Dioxide
Nitric Oxide
Nitrogen Dioxide
Chlorine
Hydrogen
Hydrogen Cyanide
Hydrogen Chloride
Ethylene
300 ppm
15 ppm
5 ppm
35 ppm
5 ppm
1 ppm
100 ppm
10 ppm
5 ppm
100 ppm
0 ppm
-2 ppm
0 ppm
0 ppm
3.5 ppm
0.5-1 ppm
0 ppm
0 ppm
0 ppm
0 ppm