Sulfur Dioxide Monitor Z-1300XP
|
|
Model Z-1300XP
Sulfur Dioxide Monitor |
Download PDF |
Theory of Operation
The sensing element of the instrument is an electrochemical cell. The cell is a four-electrode type, which contains a working and an active auxiliary electrode. The signal from the auxiliary electrode is used for temperature compensation and to improve the selectivity of the entire sensor. The sensor response is linear with the concentration of the target gas in air. |
Interferences |
Gas |
Concentration |
Meter Output |
Carbon Monoxide |
300 ppm |
3 ppm |
Hydrogen Sulffide |
15 ppm |
20 ppm |
Nitric Oxide |
35 ppm |
-1<x<0ppm |
Nitrogen Dioxide |
5 ppm |
-6 ppm |
Chlorine |
1 ppm |
-0.5<x<0ppm |
Hydrogen |
100 ppm |
0 ppm |
Hydrogen Cyanide |
10 ppm |
5 ppm |
Hydrogen Chloride |
5 ppm |
1 ppm |
Ethylene |
100 ppm |
0 ppm |
|
|
Features
- Portable
- Internal Pump
- Rechargeable Battery
- Remote Sampling
- Simple to Use
- TWA, STEL and Peak
- Audible Adjustable Alarm
- Data Logging
|
Description
Environmental Sensors Co.'s Sulfur Dioxide Meter is a portable desktop instrument that measures sulfur dioxide concentration in a range of 0-20 ppm and a resolution of 0.1 ppm.
The instrument makes it possible to monitor sulfur dioxide vapor in air. The insturment has an LCD display giving concentrations in ppm every 10 seconds, a low battery indicator, and an audible alarm that can be set at any level from 0-20 ppm.
With the touch of a button, the meter displays STEL (average of every 15 min.), TWA (8 hour average) and Peak.
The internal pump makes it possible to sample from hard-to-reach areas or confined spaces by attaching a flexible tube to the meter.
Data Logging
The Z-1300XP meter stores all of the exposure points for up to 5 logs (a log is created in the instrument's internal memory each time it is used). A log contains: date, time, number of exposure points, the interval between each point, and the actual exposure points. All of the log files can be easily uploaded to PC using components vailable within the Microsoft Windows Operation System or the termianl software included with the instrument. |
Meter Spec |
Sensor Type |
Electrochemical |
Nominal Range |
0-20 ppm |
Maximum |
100 ppm |
Sensor Life |
> 2 years |
Resolution |
0.1 ppm |
Min. Detectable |
0.1 ppm |
Response Time |
<15 sec. |
Warm-Up Time |
60 sec. |
Operating Temperature |
-20 °C to 50 °C |
Pressure Range |
Atm ± 10% |
Relative Humidity
|
15 - 90 % non-condensing |
Alarm |
Audible 80db |
Dimensions |
H x D x W: 7.5" x 5.75" x 2.75" |
Weight |
900 grams |
Battery |
Internal Rechargeable Battery or AC Adaptor |
Warranty |
One Year |
|
|
|
|